Perchlorate

ClO₄⁻

The ELIT perchlorate electrode operates over a very wide pH range and offers excellent selectivity over common anions such as chloride and sulphate. Used in water monitoring and industrial analysis.

Anion (−)ELIT 8061Valence: Monovalent
Electrode Specifications
Model
ELIT 8061
Membrane Type
Solid-state PVC polymer matrix membrane
Ion Type
Monovalent anion
Molar Mass
99.451 g/mol — 1000 ppm = 0.010 M
Physical Size
130 mm body (excl. contact) × 8 mm diameter. DC resistance < 2.5 MΩ at 25°C. Min. sample volume: 5 ml.
Operational Specifications
Concentration Range
0.2 to 9,960 ppm (2×10⁻⁶ to 0.1 Molar)
Electrode Slope at 25°C
54 ± 5 mV/decade
pH Range
pH 0 to 11
Temperature Range
0 to 50°C
Response Time
< 10 seconds (90% response)
Potential Drift
< 3 mV/day in 1000 ppm (8 hours)
Reagents
Reference Electrode
Single junction (ELIT 001).
ISAB / Buffer
1M CH₃COONa (sodium acetate) — Add 2% v/v.
Standard Preparation
Dissolve 1.231g anhydrous sodium perchlorate (NaClO₄) in 1 litre deionised water.
Interference

The Selectivity Coefficient (SC) represents the approximate apparent increase in the measured concentration caused by 1 unit of the interferent. Effect (% error) = (interferent concentration × SC / analyte concentration) × 100.

Interfering IonSelectivity CoefficientNotes
Thiocyanate (SCN⁻)0.03
Iodide (I⁻)0.02Tolerated up to the same concentration as perchlorate (~2% enhancement).
Nitrate (NO₃⁻)0.02Tolerated up to the same concentration as perchlorate (~2% enhancement).
Chloride (Cl⁻)0.0003Can be present up to 100× more concentrated before significant error.
Phosphate (PO₄³⁻)0.0002
Acetate (CH₃COO⁻)0.0001
Standard Analytical Method

Apparatus Required

  • Ion-Selective Electrode: ELIT 8061
  • Reference electrode: Single junction (ELIT 001)
  • Dual electrode head (ELIT 201)
  • ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter
  • 150 ml polypropylene beakers, 100 ml volumetric flask, 1, 2, 5, 10 ml pipettes

Calibration

Calibrate with 1000, 100, 10, 1, 0.1 ppm ClO₄ solutions. For high ionic strength samples (> 0.01 M), add 2 ml ISAB to each 100 ml standard.

Sample Preparation & Measurement

Low ionic strength: immerse in 50–100 ml sample. High ionic strength: add 2 ml ISAB to 100 ml sample and stir well.

Results

Results are displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples, figures will not need adjusting as all are affected by the same dilution factor. Allow 2–3 minutes stabilisation after electrode immersion. Wash and dry electrodes between samples to avoid cross-contamination.

Analytical Notes
  • Wide pH operating range (0 to 11).
  • Nitrate can be tolerated at equal concentration to perchlorate (causes ~2% enhancement).
  • Chloride can be present up to 100× the perchlorate concentration before causing significant error.