Barium
Ba²⁺The ELIT barium electrode requires a unique still-solution measurement protocol and Standard Addition for samples with high ionic strength, due to interference from common ISAB constituents.
The Selectivity Coefficient (SC) represents the approximate apparent increase in the measured concentration caused by 1 unit of the interferent. Effect (% error) = (interferent concentration × SC / analyte concentration) × 100.
| Interfering Ion | Selectivity Coefficient | Notes |
|---|---|---|
| Strontium (Sr²⁺) | 0.09 | Highest interferent — unlikely in most samples. |
| Potassium (K⁺) | 0.02 | Significant error if K > 10× Ba concentration. |
| Sodium (Na⁺) | 0.02 | Significant error if Na > 10× Ba concentration. |
| Magnesium (Mg²⁺) | 0.006 | Tolerated up to ~20× Ba concentration. |
| Ammonium (NH₄⁺) | 0.003 | — |
| Calcium (Ca²⁺) | 0.002 | Tolerated up to ~50× Ba concentration. |
| Lithium (Li⁺) | 0.002 | Tolerated up to ~50× Ba concentration. |
Apparatus Required
- Ion-Selective Electrode: ELIT 8081
- Reference electrode: Single junction AgCl (ELIT 001)
- Dual electrode head (ELIT 201)
- ELIT Computer Interface/Ion Analyser, or Ion/pH/mV meter
- 150 ml polypropylene beakers, 100 ml volumetric flask, 1, 2, 5, 10 ml pipettes
Calibration
For low ionic strength samples, use Direct Potentiometry with full calibration (1, 10, 100, 1000 ppm Ba). For high ionic strength, use Standard Addition: calibrate to find the slope, then enter slope and expected concentration into the software.
Sample Preparation & Measurement
Measurements must be made in still (unstirred) solutions. After immersion, manually swirl beaker for ~30 seconds, then leave to stand. Record mV at the lowest point before values start to rise again (typically < 2 mins for 10 ppm and above; up to 5 mins for < 1 ppm).
Results
Results are displayed as ppm and mol/l. If buffer solution has been added equally to standards and samples, figures will not need adjusting as all are affected by the same dilution factor. Allow 2–3 minutes stabilisation after electrode immersion. Wash and dry electrodes between samples to avoid cross-contamination.
- Unlike most ISEs, barium MUST be measured in still (not stirred) solutions.
- No ISAB is recommended — common ISAB constituents (K, Na, NH₄) interfere with the barium electrode.
- Divalent cation — slope (~21 mV/decade) is slightly lower than a pure Nernstian response for a divalent ion.