Elyxir Solutions · Nico2000 Ltd

ELIT-ORP-31C

ORP Combination Electrode — General Applications
Product Datasheet
Platinum-Rod Sensing Element
Gel-filled · Plastic Body
Ag/AgCl Reference

Overview

The ELIT-ORP-31C is designed for general ORP (Oxidation-Reduction Potential) measurements in laboratory, field, and light industrial environments where the electrode is not submerged or operated under pressure. Its rugged gel-filled plastic body and integrated Ag/AgCl reference make it straightforward to use with any BNC-equipped meter or analyser.

Specifications

ParameterValue
Body Dimensions12 × 120 mm
Body MaterialPlastic, gel-filled (sealed)
Sensing ElementPlatinum rod
Reference SystemAg/AgCl in saturated KCl gel
Reference Offset~200 mV vs Standard Hydrogen Electrode
Eh = Emeasured + ≈200 mV
Cable / Connector1 m (optional 3 m), BNC
Temperature Range0 °C to 80 °C
Protective SkirtYes
Temperature CompensationNone (ORP measurements are temperature-dependent)

Applications

Aquaria Aquaculture Swimming Pools Ponds Water Quality Disinfection Corrosion Measurement Cooling Towers Pollution Control Pulp Bleaching Wastewater Treatment

Preparation & Operating Instructions

  1. Remove the protective storage bottle and keep it for future storage.
  2. Rinse any salt deposits from the electrode tip with distilled water.
  3. Connect the BNC connector to your meter or analyser.
  4. Allow a short stabilisation period before recording measurements.
ORP measurements are temperature-dependent. No automatic temperature compensation is available — record sample temperature alongside readings where accuracy is critical.

Understanding ORP Values

ORP (Oxidation-Reduction Potential) indicates the oxidising or reducing capacity of a solution. A higher reading indicates more oxidising agents present.

ApplicationTypical ORP Range
Freshwater aquarium250–300 mV
Saltwater aquarium350–400 mV
Aquarium — harmful for fish> 400 mV
Pond — good water quality~300 mV
Pond — decomposing organics< 150 mV
Pond — chlorine / cleaning agents> 450 mV
Swimming pool (ideal, pH 7.2)≥ 700 mV

Maintenance & Storage

  1. Storage: Keep the electrode in the protective storage bottle filled with saturated KCl solution when not in use.
  2. Cleaning:
    • Non-organic contamination: soak in 0.1 M HCl for 10 minutes.
    • Organic / oily contamination: soak in 1:10 diluted detergent for 10 minutes.
    • After any chemical cleaning: re-soak in saturated KCl for 3 hours.
  3. Reference clogging: The Ag/AgCl reference may become clogged when used in samples containing TRIS buffer, proteins, or sulfides.

Performance Verification

Test SolutionExpected Response
pH 7 buffer + Quinhydrone86 ± 15 mV
pH 4 buffer + Quinhydrone263 ± 15 mV
ZoBell's solution (25 °C)238 mV